Enteral / Nutrition
Enteral Feeding Connector
Enteral feeding connectors are governed by ISO 80369-3 (ENFit) to prevent dangerous misconnection with IV luers. The material set needs tight dimensional control at the connector, lipid and nutrition-formula resistance, and soft, kink-resistant tubing for prolonged contact.
Sterilization: GammaE-BeamEthylene Oxide
Component material package
A recommended grade per component, drawn from the Nexeo catalog. Recommendations are engineering assessments; a cited property links the choice to data we hold for that grade.
| Component | Recommended grade | Why |
|---|---|---|
ENFit connector Misconnection-proof rigid fitting | Purell RP374R Polypropylene (PP) | A random-copolymer PP holds the ISO 80369-3 thread geometry and resists the lipids in enteral formula. |
Feeding tube Soft prolonged-contact conduit | Tecoflex EG-80A Thermoplastic Polyurethane (TPU) | An aliphatic TPU gives soft, kink-resistant, non-yellowing tubing suited to prolonged enteral contact. |
Strain-relief boot Flexible connector-to-tube transition | THERMOLAST M TM5MED Thermoplastic Elastomer (TPE) | A medical TPE overmolds the connector-to-tube junction for strain relief without adhesives. |
Producer datasheetNexeo Applications Engineering assessmentAwaiting datasheetNot published by producer
Key regulatory requirements
- ISO 80369-3 (ENFit) enteral connectors
- ISO 10993-1 prolonged contact
- DEHP-free / non-PVC increasingly specified for enteral
Engineering watch-outs
- ENFit geometry is tightly toleranced — qualify the connector grade for dimensional stability and shrink.
- Lipid resistance matters: enteral formulas are fat-rich and can swell some elastomers.
- Aliphatic vs aromatic TPU affects yellowing; confirm the grade family against the color requirement.
Adapt this package to your part
Hand this material package to the AI material advisor and adjust it for your process, contact duration, and material restrictions.