An oxygenator housing contains the hollow-fiber bundle and the blood/gas flow paths of a perfusion circuit. It is a prolonged blood-contact device that must be optically clear for clot inspection, rigid enough to hold pressure, and bondable to potting compounds.

Sterilization: Ethylene OxideGamma
Adapt this with AI

Component material package

A recommended grade per component, drawn from the Nexeo catalog. Recommendations are engineering assessments; a cited property links the choice to data we hold for that grade.

ComponentRecommended gradeWhy
Main housing shell
Clear pressure-holding enclosure
Makrolon Rx2530
Polycarbonate (PC)
A high-flow clear PC fills large thin-wall housings while keeping the clarity needed to inspect the blood path. Grade selected for melt flow and impact; clarity is a family characteristic not yet on this grade sheet.
End caps / manifolds
Flow-distribution end caps
Makrolon Rx1805
Polycarbonate (PC)
A medical PC matches the shell for solvent-bonding and holds connector geometry.
Inlet/outlet connectors
Rigid barbed tubing connectors
Eastman Tritan MX711
Copolyester (Copolyester)
A clear copolyester resists the lipids and alcohols in a perfusion circuit and bonds to PC assemblies.
Producer datasheetNexeo Applications Engineering assessmentAwaiting datasheetNot published by producer

Key regulatory requirements

  • ISO 10993-4 hemocompatibility (blood contact)
  • ISO 10993-1 prolonged contact
  • USP Class VI
  • Burst-pressure and connector pull per applicable device standard

Engineering watch-outs

  • Blood contact requires hemocompatibility (ISO 10993-4), which is beyond a Class VI statement.
  • Solvent bonding PC introduces residual-solvent and stress-crazing risk — validate bond process.
  • Confirm the potting-compound-to-housing adhesion as a system, not per material.

Adapt this package to your part